Use of dental implanter and maintenance of implant motor



Dental Implant Motor Features:

1) Intuitive LCD display and control panel bring convenience to the operation;

2) Original imported brushless motor is characterized by great torque, low noise, non-vibrations as well as accuracy;

3) Alternative input supply voltage: AC100-110V or AC220-240V;

4) Multifunctional foot switch is not only handy to operate but also against cross-infection caused by the contact between hands and the device;

5) Adjustable rotation speed, torque, coolant flow and forward/reverse, which can meet various demands during implanting process;

6) 10 types of reduction contra angles are able to be selected in order to meet various clinical demands;

7) 10 kinds of program settings. Each of which allows setting up different values ​​of rotation speed, torque, coolant flow, forward/ reverse and reduction ratio.

Specification:

Power Supply: AC100-110V or AC220-240V 50/60HZ

Conductivity Detection Range: 210W

Motor Rotation Speed ​​Range :1000rpm-30000rpm (1:1)

Torque Range: 5.0-55N.cm (20:1)

Coolant Flow: 0-100ml/min

Stored Programs: 10 types

Foot Switch: Control coolant flow, program forward/reverse, speed





Maintenance of control unit and foot switch:

When wiping off the blood or normal saline on the control unit and the foot switch, please unplug the power cord first, wipe it off with a cloth soaked in water (after wringing it out), and then carefully wipe it with a cloth containing disinfectant alcohol go.

Automatic protection circuit:

When the motor is overloaded, in order to protect the motor and the control host, the protection circuit starts to activate, and the power supply to the motor is automatically cut off.

How to release the protection circuit

Release and then step on the foot switch to release the action of the protection circuit.

Replace the fuse:

If the voltage is abnormal, plug in the power supply and start the machine, but the machine does not respond, please check the fuse.

How to replace the fuse? Please use a tool to pry open the fuse box on the back of the control host, pull out the old fuse, and replace it with a new one. 

How to maintain the dental handpiece:

After each treatment, rotate the handpiece of the handpiece in clean water for 4-5 seconds to wash off the attached blood and saline. If there is too much dirt attached to the body of the dental handpiece, wash it with water and wipe it off with a dry soft cloth. Do not fully immerse the handpiece in water to prevent water from entering the handpiece from the rear of the handpiece.

Specific operations:

After each operation,

1. Wipe the surface of the dental handpiece with 75% alcohol gauze.

2. Then replace the salt water with distilled water, and run the motor to rinse the inside of the dental handpiece.

3. Disassemble the dental handpiece and soak it in a special disinfectant solution for 10~15min. Alcohol and hydrogen peroxide are prohibited.

4. Scrub the blood stains with a soft brush, place it in an ultrasonic vibration cleaning machine, pay attention to separating it with a rubber pad, and finally ultrasonically clean it with distilled water
5. Oil injection lubrication
Lubricate the thoroughly cleaned and dry parts separately for assembly.
Use of dental implant: used for perforation of jaw bone or implant drilling during oral implant treatment.

Indications for dental implants:

1.Patients with missing one tooth, multiple teeth, or all teeth.
2.The alveolus of the mandibular full mouth is completely absorbed, it is difficult to wear traditional full mouth dentures, the function is not good, and it is difficult to wear them securely.
3.Those who wear traditional removable dentures with poor fixation and cannot tolerate non-functional mucous membranes.
4.Those who have special requirements for the appearance and function of dentures.
5.There is a problem with the remaining teeth around the missing teeth, and traditional dentures cannot be used.

Prohibited items:

1.Be sure to replace the bur or attach and detach the dental handpiece after the rotation of the motor has completely stopped. (May cause injury.)

2.After the treatment, please clean it immediately, lubricate it with oil (do not lubricate the motor with oil), and sterilize it at high temperature and high pressure. (May cause blood to clot inside the dental handpiece.)

3.In the environment of electromagnetic waves, malfunction may occur. Do not place this product near machines that generate electromagnetic waves.