The maintenance/ repair of dental handpieces
Dental handpieces used in dental clinics are precision micro-equipment. In use, in addition to daily disinfection of dental handpieces, correct use and maintenance are important measures for the normal work and prolongation of the handpiece's life; incorrect use will cause work Low speed, high noise and vibration, broken bearing ball frame, etc., greatly reduce the service life.
★ Proper use of dental handpieces ★
1 before use
(1) Before installing the dental handpiece, press the foot switch to spray out the remaining air and water in the tube, and observe whether the air and water are normal;
(2) install dental handpiece bur;
(a) Screw type: Install the dental handpiece on the treatment table, install the bur, and use the dental handpiece key to tighten it gently. Don't use too much force, just put it in place.
(b) Button type (gland type): press the button on the head of the dental handpiece to load and unload the bur, release the button, and the bur will be clamped. After installing the bur, check whether it is qualified by hand.
(3) Jog the foot switch to adjust the water to an appropriate flow rate and the air to an appropriate pressure. The standard pressure of the high-speed dental handpiece is between 0.2-0.25kgf/cm2.
( 4 ) Note: Before the dental handpiece is equipped with a bur, it is strictly forbidden to step on the foot switch, otherwise the three-leaf spring will rotate out, which will push against the back cover and both will be damaged.
2 in use
(1) When using the dental handpiece, it must be careful, meticulous, and light. It is strictly forbidden to operate brutally, and do not drop the dental handpiece to the ground.
(2) No matter what the angle is, it should be operated according to the way of lifting up and down, so as to achieve light touch and jog, and never use force from side to side.
(3) Use the improved pen grip to hold the dental handpiece, find a stable fulcrum, and perform intermittent operations.
(4) According to the different tooth positions and the different shapes of the prepared holes, replace the dental handpiece and bur in time to minimize the wear on the bearing of the dental handpiece.
(5) Sharp burs should be used. Blunt burs are not only time-consuming and labor-intensive, but also cause damage to the dental handpiece and increase the pain of the patient.
(6) During the use of the dental handpiece, always check whether the needle is loose to avoid flying burs.
(7) Choose different dental handpieces according to different clinical situations (for example, use a miniature handpiece that does not affect the line of sight for posterior tooth pulp opening; use a powerful high-torque handpiece for tooth preparation).
(8) The one-time workload is too large, and two dental handpieces should be used in turn to avoid damage to the dental handpieces due to excessive load, and to save time for replacing dental handpieces.
(9) The radial swing of the rotor of the dental handpieces is serious, and it is strictly forbidden to use it when it is moving longitudinally, and it is strictly forbidden to use the dental handpieces in an anhydrous or water-deficient state.
(10) The air pressure and water pressure will change with the use of the treatment table, so the air pressure and water pressure should be checked frequently during clinical operations to ensure the normal use of the dental handpieces.
★ Maintenance of dental handpieces ★
1 cleaning
After using the dental handpiece, it should be cleaned as soon as possible, otherwise the dirt will solidify and be difficult to clean. Before unloading the bur, clean debris from the working head with water mist, then scrub with 75% alcohol.
2 Take the bur
After use, you only need to loosen (turn) the key by 1/4 turn when removing the needle. If the key is too loose, the three-leaf spring will be pushed against the back cover and damaged. After each use, the burs should be removed in time to facilitate the recovery of the bur clamping mechanism of the dental handpiece.
3 come on
Air pressure oil spray can is the most ideal lubrication tool, it can not only lubricate the dental handpiece, but also achieve the effect of cleaning the bearing and wind wheel.
(1) Before using the oil spray tank, remove the bur (the screw-type dental handpiece can be reversed by 1/4 turn), wrap the handpiece with paper, and insert the nozzle of the oil tank into the air inlet of the dental handpiece to ensure sufficient air pressure.
(2) The air pressure oil spray tank must be used vertically, and each spray is 2-3s.
(3) For 2-hole dental handpieces, the nozzle is aimed at the large hole for spraying; for 4-hole dental handpieces, the nozzle is sprayed for the second largest hole.
(4) If it is oil dripping method, drop 4 drops of oil into the large hole, and then use an air gun to blow air intermittently into the oil filling hole for 30 seconds.
(5) It is strictly forbidden to use industrial lubricating oil for lubrication to prevent patients from being poisoned.
(6) Dental handpieces that are not used for a long time should be filled with oil, sealed and stored.
★ Repair of dental handpiece ★
Dental high-speed handpieces come in a wide variety. According to the up and down needle method of the bur, it can be divided into rotary handpiece, push-type handpiece, and push-type handpiece; according to the size of the handpiece, it can be divided into mini-type handpiece, standard handpiece, and torque-type handpiece; according to whether it is equipped with optical fiber It can be divided into handpieces with optical fiber and handpieces without optical fiber.
To overhaul a machine, you must first understand the basic principles and usage of the instrument. Without knowing how to use the instrument correctly, it is difficult to recognize the abnormal state of the instrument. In this case, it is very difficult, sometimes even impossible, to do a good job of overhauling the instrument. The same goes for servicing a dental handpiece, which should begin with a "ask," "see" and "check" approach.
Question: Refers to asking the user about the process of failure, such as whether the head of the dental handpiece does not come out of water, whether it is rotating, whether there is noise (rustling, howling), and whether it is shaking;
Look: just look at the appearance of the dental handpiece, if it has been bumped and dropped, there must be traces of the collision on the appearance;
Inspection: It is to install a standard rod or bur and turn it by hand, listen to the sound or feel whether there is any obvious abnormal phenomenon inside, and then carefully check and prescribe the right medicine, which is very important for developing correct maintenance ideas.
Maintenance is to find out the cause according to the fault phenomenon, eliminate the fault, provide technical support for the doctor, and treat the specific reason in detail.
Fault 1: Poor water supply occurs
Including: (1) Insufficient water supply; (2) Uneven water volume; (3) Water supply is not foggy; (4) No water at all.
Check where the water is blocked:
( 1) Tubing: Remove the turbo dental handpiece and check if a proper amount of water flows out of the tubing. If no water comes out, there may be a problem with the plumbing or the unit's water supply system (poor connection or malfunction, and insufficient water pressure at the water source)
( 2) Interface: Connect the interface with wires for trial operation. If the proper amount of water does not flow out, it indicates a problem at the connection (limescale, etc. clogs the water supply system or the connection is loose, causing insufficient pressure).
(3) Dental handpiece body: After removing the shaft core from the turbine, install the interface for operation. If the proper amount of water is not coming out, then there is a problem within the turbo handpiece body (dirt in the water supply hole or blockage of the water supply in the turbo handpiece).
(4) Shaft core: If the water supply in the above 3 items flows out normally, it is the problem of the shaft core (the water supply nozzle is blocked).
Fault 2: The nozzle of the shaft core is blocked
Solution:
( 1) Remove the end cover and take out the shaft core.
(2) Remove the nozzle cleaning needle behind the brush, and insert the needle straight into the nozzle.
(3) The nozzle cleaning needle should be inserted to the upper groove (the needle passes through the air injection groove). Use a special brush to clean this part of the dust.
(4) Blow off with a three-purpose gun, etc. from the bur side.
(5) When installing the shaft core on the dental handpiece head, be sure to install the O-ring seal correctly.
Fault 3: The bur is easy to fall off, the bur cannot be pulled out, and the axis vibrates
Solution:
(1) When selecting a bur, you must confirm the allowable speed range specified by the bur manufacturer, otherwise the bur will easily fall off when the cutting load is greater than the clamping force of the chuck.
(2) The short-handle bur should be used for the mini dental handpiece head, and the long-handle bur is easy to fall off when it is installed.
(3) Do not use worn and thinned burs, bent burs, or burs with missing edges, otherwise the durability of the bearing will be affected.
(4) After the bur is installed, the bur must be pulled out to confirm that it is firmly installed. If it is not inserted deep, or too shallow, it will cause the danger of the bur flying out.
Fault four: The dental handpiece does not turn or the speed is slow
Solution:
(1) The head of the dental handpiece collides, falls and deforms, which hinders the operation of the wind wheel. As long as it is restored with the right tool, the wind wheel can be made to operate flexibly.
(2) Bearing lacks oil lubrication. Lubricate with dental handpiece lubricant.
(3) The bearing is damaged. Just replace the bearing with the same type.
(4) Low air pressure. If it cannot reach 2. 0 - 2. 5kgf /cm2, adjust the air pressure to stabilize the air pressure at 2. 0 - 2. 5kgf /cm2, and the speed will be normal
Maintenance Note:
It is forbidden to use burs with non-standard handles
Do not use bent or deformed burs
Do not press the cover until the dental handpiece is completely stopped
It is forbidden to immerse the dry heat disinfection box in chemical solution for disinfection, and it is recommended to use high temperature and high pressure steam for disinfection
Do not use dental handpieces without disinfection
Unsterilized dental handpieces are prohibited from being sent for repair
Good habits to prolong the life of dental handpieces:
1. The sand produced by the sandblasting machine without "anti-suckback" is very harmful to the bearing of the dental handpiece, so try not to use it in the same connecting pipe
2. Ultrasonic cleaners can cause great damage to bearings, so do not use ultrasonic waves to clean dental handpieces
3. The clamping system of the dental handpiece is not durable, or the bur cannot be clamped. Check whether the handle does not meet the standard bur, or the use of inferior burs has damaged the movement of the dental handpiece.
4. Do not increase the pressure by yourself because the drill is not moving. In most cases, the bur is worn out. It is recommended to replace the bur
5. The compressor needs to be cleaned, filtered and waterproofed frequently, otherwise the water in the air will damage the bearing.
Tuesday, February 7, 2023
Thursday, August 18, 2022
The structure of the dental handpiece
Dental handpieces are classified into two types: ball bearing type and air bearing type, according to the different working methods of the bearings used.
Ball bearing type high-speed handpiece is mainly composed of handpiece, handle and dental handpiece connector.
1. The internal structure of the machine head
The vortex core rotor is the core component of the machine head. It is composed of a bearing, a wind wheel and a mandrel. There are upper and lower miniature bearings in the machine head. The bearing is cushioned with an O-shaped rubber ring to keep the bur at high speed. Radial and axial concentricity during rotation, and reduces operating noise and vibration.
2. Clamping method of bur
There are various ways for the machine head to clamp the bur. In the early days, the reed-type clamp was used, that is, there were 3 evenly divided bow-shaped reeds in the mandrel. The bur handle is inserted into it, and the reed is deformed by force to clamp the needle handle.
Nowadays, screw-type and pressing-type bur clamping methods are mostly used. The clamping sleeve of this dental handpiece has an elastic three-lobed clamping spring. The clamping spring is in the shape of a hammer and has three axial grooves. Or press the top spring to open or lock the clamp spring to hold the bur.
3. Spray structure
There is a water pipe and a gas pipe at the tail of the turbine dental handpiece. The water and gas are mixed at the outlet of the handpiece and sprayed out from the spray hole in the form of a mist.
4. Dental handpiece connector
The dental handpiece connector is the connection between the handpiece and the output hose of the main unit. The main air flow that drives the rotation of the dental handpiece wind wheel and the branch air that generates atomization enter the main air hole, branch hole and water hole of the dental handpiece connector through the pipeline respectively. Dental handpiece. There are two ways of connection between the handpiece and the handpiece connector: one is with a fixed lock nut on the handpiece connector, and the other is with a quick-release connector.
RUITO dental equipment company offer all kinds of dental handpiece with good price,if want to know more,please check the website:www.ruitodental.com.
Monday, May 23, 2022
Use of dental implanter and maintenance of implant motor
Dental Implant Motor Features:
1) Intuitive LCD display and control panel bring convenience to the operation;
2) Original imported brushless motor is characterized by great torque, low noise, non-vibrations as well as accuracy;
3) Alternative input supply voltage: AC100-110V or AC220-240V;
4) Multifunctional foot switch is not only handy to operate but also against cross-infection caused by the contact between hands and the device;
5) Adjustable rotation speed, torque, coolant flow and forward/reverse, which can meet various demands during implanting process;
6) 10 types of reduction contra angles are able to be selected in order to meet various clinical demands;
7) 10 kinds of program settings. Each of which allows setting up different values of rotation speed, torque, coolant flow, forward/ reverse and reduction ratio.
Specification:
Power Supply: AC100-110V or AC220-240V 50/60HZ
Conductivity Detection Range: 210W
Motor Rotation Speed Range :1000rpm-30000rpm (1:1)
Torque Range: 5.0-55N.cm (20:1)
Coolant Flow: 0-100ml/min
Stored Programs: 10 types
Foot Switch: Control coolant flow, program forward/reverse, speed
Maintenance of control unit and foot switch:
When wiping off the blood or normal saline on the control unit and the foot switch, please unplug the power cord first, wipe it off with a cloth soaked in water (after wringing it out), and then carefully wipe it with a cloth containing disinfectant alcohol go.
Automatic protection circuit:
When the motor is overloaded, in order to protect the motor and the control host, the protection circuit starts to activate, and the power supply to the motor is automatically cut off.
How to release the protection circuit
Release and then step on the foot switch to release the action of the protection circuit.
Replace the fuse:
If the voltage is abnormal, plug in the power supply and start the machine, but the machine does not respond, please check the fuse.
How to replace the fuse? Please use a tool to pry open the fuse box on the back of the control host, pull out the old fuse, and replace it with a new one.
How to maintain the dental handpiece:
After each treatment, rotate the handpiece of the handpiece in clean water for 4-5 seconds to wash off the attached blood and saline. If there is too much dirt attached to the body of the dental handpiece, wash it with water and wipe it off with a dry soft cloth. Do not fully immerse the handpiece in water to prevent water from entering the handpiece from the rear of the handpiece.
Specific operations:
After each operation,
1. Wipe the surface of the dental handpiece with 75% alcohol gauze.
2. Then replace the salt water with distilled water, and run the motor to rinse the inside of the dental handpiece.
3. Disassemble the dental handpiece and soak it in a special disinfectant solution for 10~15min. Alcohol and hydrogen peroxide are prohibited.
Dental endo motor with apex locator operate video
Feature
Six Functions;
German Motor;
Large Colorful OLED Screen;
Four Working Modes;
R-Smart Plus New Apex Locator with Endo Motor.
Technical Data
Battery: 7.2V/1400mAh
Adapter: 100V~240V 0.8A 50Hz/60Hz
Revolving speed: 200~600rpm
Torsion: 0.6~4.0N.cm
Dimensions: 250mm*210mm*140mm
Buzzer alert: The buzzer will alert when the file is less than 2mm closed to the apex
www.ruitodental.com
Email:ruito@ruitodental.com
Tuesday, May 17, 2022
How to Find The Best Dental Equipment Dental Handpiece
Monday, May 16, 2022
Here's a video on how to install a dental X-ray machine sensor.
Monday, April 18, 2022
Maintenance of Dental Autoclave Sterilizer
1.1 Scrub the exterior of the sterilizer and keep the sterilization chamber clean and dry every day
1.2 Cleaning of the filter screen, the filter screen is easily attached to various fiber debris and sediments, which will seriously affect the speed of vacuuming and the smoothness of condensed water treatment
1.3 Check whether the rubber ring is damaged or not and whether the air inlet is blocked. It is best to apply talcum powder on the rubber strip after using it every day to prolong the service life of the rubber strip.
1.4 Check whether the pointer of the pressure gauge can return to "0", and the degree is not accurate
1.5 After the sterilization work is over, cut off the power supply, turn off the power supply, and turn off the steam power supply.
2.1 Clean the inner wall of the sterilization pot with neutral detergent every week; remove the remaining water inside the boiler and close the valve in time
2.2 Pull the handle of the safety valve in the inner chamber of the sterilizer once a month and flush it with steam to prevent its action from malfunctioning. At the same time, it should be noted that the sterilization room must be thoroughly descaled, repaired and maintained at least once a month. When descaling, properly mix water and descaling agent according to the principle of 1:100, and then perform descaling after mixing evenly. After 10 minutes, clean the container and then wipe the container with a lint-free cloth to keep the container dry. The sterilization room and the cleaning car can also be cleaned according to the above method, and the stainless steel parts should be wiped with gauze soaked in paraffin oil.
2.3 Every quarter, let the pressure sterilizer manufacturer carry out large-scale maintenance and repair work on the instrument. For example, check whether the door panel of the autoclave is flat, and remove the cover panel in the autoclave. At the same time, rust should be removed for the rusted parts, and high temperature lubricating oil should be added to the pressure sterilizer.
2.4 Remove the sealing strip every half a year, clean it with soapy liquid, wipe it with alcohol and dry it, and then put it back into the sealing tank; descale the steam generator; clean the air inlet and water inlet pipe filters; And steam generator pressure gauge sent to the Measurement Technology Testing Institute for testing and retain the test certificate.
2.5 The sterilizer and the safety valve of the steam generator are sent to the Special Equipment Supervision and Inspection Institute for inspection once a year and the inspection report is retained.
Thursday, March 24, 2022
Correct use of transparent scaler
Periodontal disease is one of the frequently-occurring and common diseases in oral diseases. The statistical incidence rate is high in all countries in the world. Teeth cleaning is an ideal treatment method for periodontal disease. Ultrasonic scaler is an efficient scaler. It uses high-frequency vibration energy to remove calculus and plaque, especially for calculus, soft scale and pigment attached to the tooth surface. The rational use of ultrasonic scaler is very important. Improper operation will cause damage to the tooth and even the pulp tissue, and will also cause premature wear of the cutter head and loss of power.
1. Principle of action
The ultrasonic scaler consists of two parts: an ultrasonic generator and a converter, and is equipped with a mobile phone, a replaceable working head, and a foot switch. Its working principle is to convert electrical energy into high-frequency mechanical energy, so that the working head vibrates at a high speed. The frequency is between 25kHz and 32kHz. This high-frequency vibration can break the calculus and cause it to fall off the tooth surface. Since the ultrasonic wave causes the high-frequency vibration of the working head to generate a lot of heat, it must be equipped with a cooling system. At the same time, through the cavitation of water It generates pressure locally to clean the surgical area.
2. Clinical Operation
1). Use the modified pen type to hold the handpiece of the instrument, and make a conventional fulcrum in the mouth to stabilize the instrument.
2). Make the long axis of the working head parallel to the long axis of the tooth body, the working angle is not more than 15°, and try to make the working head fit the tooth surface as much as possible.
3). The working head should not stop on the tooth surface. Always keep it moving on the tooth surface. The end of the working head cannot be vertically aligned with the tooth surface. When removing calculus, the pressure should be light, and the calculus should be dislocated by vertical or oblique movements. If the working head is stuck between the teeth, the work should be stopped immediately. Then gently withdraw in the direction of the abduction gap, do not twist strongly, and do not try to withdraw while maintaining the working state, otherwise the cutter head will be easily broken, especially the powerful imported scaler.
4). When the patient wears a porcelain crown and bridge, and the adjacent occlusal surface has fillings, photosensitive resin and other restorations, the working head should not move directly on the coronal surface or fillings, so as to avoid crown cracks or fillings falling off.
5). During subgingival scaling, the side of the working tip moves gently back and forth along the gingival margin, and gradually moves down. The end of the working tip is long and tapered, which can extend into the periodontal pocket, and insert the probe into the periodontal pocket. To a certain depth, guide the working tip to move slowly along the incisal plane of the tooth for 30-60 seconds, and then remove the working tip from the periodontal pocket.
3. Performance characteristics
1). Ultrasonic scaler has the advantages of high speed, high efficiency, time saving and labor saving, constant vibration frequency, high strength, strength, durability, easy removal of calculus, small postoperative reaction of patients, and light tooth sensitivity.
2). Water volume adjustment, the working head is in the form of water mist as the standard. When there is dripping water, it means that the water volume is too large, and the water volume is too small, which will burn the patient's mucous membranes, lips, and damage the teeth and machines.
3). Ultrasonic vibration will produce a real cavitation effect, which is more effective in killing anaerobic bacteria in the periodontal pocket.
4). It can meet various clinical needs and has a wide range of uses, such as removing crowns, removing studs, and replacing special cutter heads for root canal treatment.
4. Cleaning, maintenance and disinfection
The shell, transformer, foot flat wrench, mobile phone line, etc. can be wiped with a neutral and clean soft cloth containing alcohol and no abrasive. The working handle can be sterilized by high temperature, high pressure and chemical disinfection. The front end of the handle can be removed for disinfection.
5. Precautions
1). When the handle is connected to the handle wire, just insert it lightly and do not turn it.
2). The shape, gloss and surface smoothness of each working tip are of extremely high standards. Please do not grind or modify it by yourself.3. When tightening and removing the working tip, it should be done in the direction of the axis of the handle.
4). During use, when the vibration is weakened, please pay attention to check whether the working tip is tightened.
5). When cleaning, just place the handle lightly on the tooth surface, no need to apply force.
6). Patients with bleeding tendency and those with an ultra-pacemaker are disabled.
www.ruitodental.com
Thursday, February 17, 2022
Common questions about using dental endo motor and clinical attention for root canal preparation
Dental endo motor Root canal instruments are widely used in oral clinics. So the question is, what are the common questions when using the Dental endo motor?
A. Can Dental endo motor be used for any root canal treatment?
R. The curved and calcified root canal hand、machine combination is safer
A. The surgeon feels that the preparation speed is too slow?
R. 250-400rpm is the safest and most efficient speed for Nitinol
A. Is the torque smaller the better?
R. It is very important to limit the torque, a reasonable torque is the guarantee of cutting force
A. Is it possible to expand while measuring?
R. The main purpose of expanding while measuring is to prevent the expansion of the apical foramen. For conditions that cannot be accurately adapted to root measuring, the expanding while measuring cannot be used.
A. Can measuring and expanding at the same time replace the root canal measurement alone?
R. Root canal measurement is affected by a variety of environments, and root canal treatment should strictly follow the sequence of "establishing a straight path - root canal measurement - preparation" to prevent accidental risks
A. What kind of machine do users use to expand needles?
R. Generally speaking, any brand of nickel-titanium expansion system can be used, each with its own characteristics, which mainly depends on the cost-effectiveness and the familiarity of the doctor.
How to prevent broken endomotor root canal file
① To prevent endomotor root canal files breakage, the operator's experience is the key
1. Learning about the machine and endomotor root canal files allows you to reduce the risk of use;
a. Control the torque and speed It is best to set the torque control motor and the matching speed. Following the torque recommended by the endomotor root canal files manufacturer, speed is especially important for treatment safety.
b. Measure the length of the instrument each time it enters the root canal. Instruments that are too long increase the risk of torsional fracture. Equipped with root canal length measuring instrument, calipers, etc. for working length measurement.
2. The correct operation method is very important: do not pressurize and force the tip. The instruments cannot stay in the root canal for too long;
3. Instruments should be checked before surgery,
a. The fracture of instruments mostly occurs in molars, especially in small and narrow root canals such as the mesial root canals of upper and lower molars, and special care should be taken before treatment;
b. Curved root canals, re-treatment and calcified root canals require special requirements for instruments and manipulations, and it is best to use a new root canal files;
4. Establish a straight path to maintain an unobstructed root canal, pay attention to gestures during operation, and try to ensure that the enlargement root canal files and the root canal path are in a straight line to reduce the bending resistance of the enlargement root canal files at the crown and the stress on the instrument;
5. The root canal should be kept fully lubricated and should be flushed every time it is prepared;
6. Because the inside of the root canal is narrow, do not put pressure on the enlargement root canal files during operation, and the cutting should be done completely by the power of the machine itself and the sharp edge of the enlargement root canal files;
7. In the case of inexperience, it is safest to use the recommended minimum torque. In general, lower rotational speeds will extend the clinical life of the device and reduce the chance of breakage.
② Efficient and safe, excellent equipment is the guarantee
1. Nickel-titanium expansion requires both rotational speed and torque. Inappropriate rotational speed and overloaded torque are important factors for root canal files breakage.
2. Control the torque and speed Follow the torque recommended by the expanding root canal files manufacturer.
3. Measure the length of the instrument each time it enters the root canal. Equipped with root canal length measuring instrument, calipers and other tools for working length measurement.
-
Dental handpieces are classified into two types: ball bearing type and air bearing type, according to the different working methods of the b...
-
Dental handpieces used in dental clinics are precision micro-equipment. In use, in addition to daily disinfection of dental handpieces, cor...